Spectrasol Applications

Sample Type: Reduction of ICP-MS
ArCl+Interference

Preparation:

Transfer 2 to 5 mls of an acid digested sample to a plastic tube.
Add 2 mls of 200 ng/mL Ge and 2 mls Spectrasol CFA-C (1+1).
Dilute to 10 mls with water.

 

Results: Comparison of detection limits (pg/mL) for As and Se isotopes with and without the use of the Spectrasol CFA-C tertiary amines reagent.

 

Analyte without 10% with 10% Signal
CFA-C CFA-C Enhancement

75As 55 6.0 9x
77Se 0.43 0.10 4.3x
78Se 0.45 0.18 2.5x
82Se 0.90 0.28 3.2x

 

Discussion:

Comparisons were done with an intermediate gas flow rate of 0.80 L/min, a nebulizer flow rate of 0.94 - 0.97 L/min and power of 1000 Watts. The 75As and 77Se isotopes which are most affected by ArCl+ exhibit the greatest signal enhancement.

 

Reference: Krushevska, A. et al. Application of tertiary amines for As and Se signal enhancement and polyatomic interference reduction in ICP-MS analysis of biological samples. Fresenius J. Anal. Chem., 355, 793 (1996).